HCL Analyzer HCL-103HCL Concentration in exhaust gas, which is discharged from buming up equipment of urban waste, is attracting the attention. HCL is generated from combustion decomposition of organ chlorine compounds such as vinyl chloride pl
HCL Analyzer is designed for measuring HCL concentration of HCL, serially and automatically, which is discharged from burning up equipment of urban waste based on the law of air pollution prevention. This analyzer executes the sampling of the fixed flow automatically, by micro computer control. In addition, the analyzer measures HCL in exhaust gas by quantifying chroline ion concentration of HCL by ion selective electrodes after absorbing HCL in the gass into the absorption liquid.
To discharge the pollutant exhausted from chimney pursuant to the standard of allowable exhaust, HCL Analyzer in TMS is used by regulating the diluted amount of pollutants in accordance with the concentration values indicated on the measuring instrument in link of nitrogen oxide and HCL. As Ministry of Environment has controlled HCL concentration values by lowering them continuously and will control the total discharge amount, higher accuracy and safety of the data values are required.
By the technology acquiring accurate and stable data values in reducing cost of waste incineration company and in transmitting data to Environment Corporation, exhaust concentration of air pollution can be reflected correctly, and also be contributed to the establishment of the policy against environmental pollution by the government.
Features of HCL
Metal film electrode is applied to detecting element, so it has better reproducibility and reliability than the devices that use the existing rod electrode or magnetic stirrer type polishing electrode.
Detecting element has temperature regulation function in its interior, so it is not affected from external temperature or chimney.
Drainage of absorbed reagent is serially executed using tube pump.
Sample gas amount is measured and auto-controlled by pressure sensor, so it is accurately absorbed by automatic digital control even in pressure changing of smoke tube.
It always operates automatic measuring of analysis liquid, and is not affected from moisture in exhaust gas.
It has high reliability of data, because checking the sensibility can be made every day by automatic sensitivity check and auto-calibration mechanism.
It can display rate value, one-hour mean value, one-day mean value, and error by micro computer control.
This is the system to make accurate measurement of electrode radical ion by removing noncontact section of electrode and liquid by bubble at unit area of measurement electrode of electrode radical ion.
This is wastewater-air treatment system that has the advantage of separating gas and liquid smoothly, since it makes gas and liquid mixed smooth by gas-liquid mixing concave section at the upper part of gas-liquid separator.
Specifications
Measurement Object
HCL gas in chimney exhaust gas
Measurement Type
24-hour sequence automatic measurement
Structure
Indoor Independent type(standar)(Option: Outdoor Independent type, Non-explosive watertight structure)
Measurement Range
0-20/50/150ppm
Measurement Method
Chloride Ion Electrode method
Repeatability
within ±2%FS(by equivalent liquid)
Zero Drift
within ±2%FS/Day(by equivalent liquid)
Span Drif
within ±2%FS/Day(by equivalent liquid)
Response Speed
Within 5 min. of 90% response (by equivalent liquid)
Sample Flow rates
500mL/min
Sample Tube Length
Standard 10m
Absorption Liquid
20L(replacement by 1 time/2weeks)
Liquid Flow Amount
Approx. 0.8mL/min
Display
Measurement value, time and measurement mode, etc.
Calibration method
2-point calibration by equivalent liquid, Manual/Auto, Remote
External Input
4-20mA DC
Contact Signal Output
Power cut-off, in regulating, malfunction, abnormal concentration and close contact output
Coated Color-Main body
Manin body:5Y7/1, CELL: Black
CPU:5Y7/1
Power
AC 110V/220V 60Hz
Electricity Consumption
400VA(Main body and Porbe are included)
Cl - Ion Electrode Method HCl Analyzer
Description
Product Inquiry
Purchase Product Catalog
Payment & Shipping Conditions:
- Payment: We accept payment through Paypal Only.
- Shipping: We will ship the catalog once the payment is received. And you will be receiving the catalog with in 10 -14 busines days. Shipping might be delayed in due to international shipping conditions which is depends on the countries receiveing. In case hard copy of catalog is not available then we will ship the soft copy.
- Refund: We can refund the order before shipping process was initiated. Incase Catalog not available, we will make sure to refund the order.
- Note: This is a Catalog Produt.
Share Cl - Ion Electrode Method HCl Analyzer in Social Media